EZ3008sc Fluoride Ion-Selective Electrode (ISE) Analyser, 1 stream, 4x mA

EZ3008sc Fluoride Ion-Selective Electrode (ISE) Analyser, 1 stream, 4x mA
# de producto: EZ3008.9700140T
Precio final Contacto Hach

Stable Performance in Changing Conditions

The EZ3008sc delivers temperature controlled, discontinuous fluoride measurements with multi stream capability. Operators get stable, reliable data during process fluctuations, reducing manual checks and adjustments while maintaining confident day to day water quality control.

Reduced Maintenance and Service Costs

Designed for long-term operation, the analyser uses gel-filled ion-selective electrodes that eliminate the need for electrolyte refilling. This technology reduces reagent consumption, routine service interventions, maximises analyser uptime, and lowers the total cost of ownership over the instrument’s lifetime.

Flexible Connectivity for Legacy and Modern Platforms

Integrate seamlessly into existing PLC, SCADA, DCS and plant network architectures with a built-in industrial panel PC. Real-time data transmission provides immediate visibility into diagnostics and status, enabling faster responses and improved decision-making across your facility.

Flexible Analysis for Multiple Streams

Conduct quick spot checks and investigations without interrupting routine monitoring using built-in grab sampling. When combined with the optional 8-channel sample stream, one analyser sequentially measures multiple process points: maximising asset utilization and coverage while simplifying your system architecture.

Contenido de la caja

EZ3008sc Fluoride Ion-Selective Electrode (ISE) Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x ISE Fluoride Electrode and 1 x empty 2.5L Reagent Container with Fittings (for Buffer Solution)

Especificaciones

Agua de refrigeración: Caudal aprox. 5 L/h; temperatura máx. 30 °C; presión máx. 0,5 bar
Agua desmineralizada: Does not apply
Aire de instrumentación: Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air. Used to flush the instrument in corrosive environement. Min. 0.2 bar - Max. 0.5 bar
Automatic cleaning: Sí; frecuencia programable a voluntad: cada 6 horas, cada 12 horas, diariamente, semanalmente
Calibración: Automatic; 2-point; Frequency freely programmable: 6 hours, 12 hours, daily, weekly. Note: manufacturer recommends that a calibration is done when the reagents are replaced
Calidad de muestra: Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Certificaciones: CE, ETL certified to UL and CSA safety standards, UKCA
Clase de protección: IP44
Conexión a tierra: Earth connection Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Contenido de la caja: EZ3008sc Fluoride Ion-Selective Electrode (ISE) Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x ISE Fluoride Electrode and 1 x empty 2.5L Reagent Container with Fittings (for Buffer Solution)
Dimensiones (A x A x P): 688 mm x 460 mm x 340 mm
Drenaje: Atmospheric pressure, vented, min. Ø 32 mm
Energía: 100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Garantía: 12 meses
Interferencias: Metal ions like aluminium [(Al)³+] > 72 mg/L, calcium [(Ca)²+] > 108 mg/L and iron [(Fe)²+/(Fe)³+] > 150 mg/L. Fats, oil, proteins, surfactants and tar.
Límite de detección Bajo (LOD): 0.05 - 2.5 mg/L F-: 0.05 mg/L
0.1 - 5 mg/L F-: 0.1 mg/L
0.2 - 10 mg/L F-: 0.2 mg/L
Material: Hinged part: Thermoform ABS, Door: PMMA
Wall section: Galvanised steel, powder coated
Método de medición: Discontinuous, direct measurement by combined Ion-Selective Electrode
  Ion-Selective Electrode
No. de corrientes de muestra: Standard: 1
Optional accessory for 2, 4, or 8 sample streams
Parámetro: Fluoride
Peso: Max. 35 kg (77 lb)
Precision: Better than 2% full scale range for standard test solutions
Presión de muestra: By external overflow vessel (open to atmospheric pressure)
Rango: 0.05 - 2.5 mg/L F-
0.1 - 5 mg/L F-
0.2 - 10 mg/L F-
Rata de flujo de la muestra: 100 - 300 mL/min
Requisitos de los reactivos: Keep between 10 - 30 °C (50 - 86 °F)
Salidas analógicas: Active 0 - 20 mA (or 4 - 20 mA) max. 500 Ohm load, standard 4, optional: 8
Salidas digitales: Relays: 5 contacts, not user configurable: Malfuntion, maintenance, analysis ready, sample ready, sample ready (EZ9150)

Ethernet Connections: Optional: Claros Ethernet connection and Modbus TCP/IP Ethernet connector; LAN version; 10/100 Mbps or Profinet or Ethernet IP

RS485 communication: Profibus DP or Modbus RTU
Temperatura Ambiente: 10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Temperatura de la muestra: 10 - 30 °C (50 - 86 °F)
Tiempo de Ciclo: Default: 5 minutes
Continuous: 5 minutes
Validación: Automatic; Frequency freely programmable: 6 hours, 12 hours, daily, weekly